Research facilities for animals (including insects) should comply with the following standard on installation and operation.

Installation Standard

Compliance Safety Management Level
1 2 3 4
Location of and Access to Laboratory Compliance with the following descriptions except when complying with Location of and Access to Laboratory in Attached Table 9-1 No.1        
Animal experiment zone: separation (division) from general experiment zone Recommended Mandatory Mandatory Mandatory
Establishment of shower facility near dressing room Recommended Recommended Recommended Recommended
Establishment of general BL3 experiment zone, such as preparation of biological product - - Mandatory Mandatory
Establishment of safe storage space for biological product - Mandatory Mandatory Mandatory
Establishment of separate space for animal transport - Recommended Mandatory Mandatory
Separation of animal breeding room and animal experiment space (surgery, dissection experiment, etc.) - Recommended Mandatory Mandatory
Installation of storage facility or space for feed, litter, etc. in animal facility - Recommended Mandatory Mandatory
Installation of exclusive washing and disinfection space for cage and apparatus and material related to animal breeding - Recommended Mandatory Mandatory
Installation of pre-disposal animal carcass storage space and treatment facility in  separate sealed zone in facility - Recommended Mandatory Mandatory
Installation of equipment for pollutant removal and back draft prevention in case of installation of drainage - Recommended Mandatory Mandatory
Experiment Zone Compliance with the following descriptions except when complying with Experiment Zone in Attached Table 9-1 No.1        
Installation of observation window with insulated glass in animal breeding room - - Mandatory Mandatory
Installation of window screen, flypaper sheet, etc. for preventing animals (including insects) from escaping - Recommended Mandatory Mandatory
Air-conditioning Compliance with the following descriptions except when complying with Air-conditioning in Attached Table 9-1 No.1        
Installation of separate airing duct - Recommended Mandatory Mandatory
Installation of HEPA filter in air supply duct - Recommended Mandatory Mandatory
Installation of carbon filter and other smell-removing equipment in exhaust system Recommended Recommended Mandatory Mandatory
Maintenance of minimum negative pressure of 70Pa in animal room - - Mandatory Mandatory
Experimenter Safety Protection Compliance with the following descriptions except when complying with Experimenter Safety Protection in Attached Table 9-1 No.1        
Preparation of motor-type respiratory protective gear fitted with HEPA filter (except for suite-type Level 4 research facility) - - Mandatory Mandatory
Experiment Apparatus Compliance with the following descriptions except when complying with Experiment Apparatus in Attached Table 9-1 No.1        
Design of cage that has lid that is not easily opened by movement of animals, etc., is easy to clean and is made from material that allows disinfection and sterilization - Mandatory Mandatory Mandatory
Establishment of facility or space, such as work cabinet with negative pressure function, for autopsy and replacement of cage, etc. - Recommended Mandatory Mandatory
Waste Treatment Compliance with Waste Treatment in Attached Table 9-1 No.1        
Other Facilities Compliance with Other Facilities in Attached Table 9-1 No.1        
Installation of pre-filter in front end of exhaust filter in animal breeding and animal experiment space (except for space using sealed-type cage - Recommended Mandatory Mandatory
Operation Standard
Compliance Safety Management Level
1 2 3 4
Access to Experiment Zone Comply with Access to Experiment Zone in Attached Table 9-1 No.2. - - - -
Activity in Experiment Zone Comply with the following descriptions except when complying with Activity in Experiment Zone in Attached Table 9-1 No.2. - - - -
Carry-in laboratory animal in individual sealed container. Recommended Mandatory Mandatory Mandatory
Transport laboratory animal in strong sealed container. - Recommended Mandatory Mandatory
Marking on genetically modified animal for identification: mark within 72 hours after birth. Mandatory Mandatory - Mandatory
Use disposable or integrated-type syringe. (After use, sterilize syringe after placing it in exclusive container before disposal.) Recommended Mandatory Mandatory Mandatory
Contaminated waste, such as culture, tissue and body fluid, and potentially infectious material: keep it in sealed container with lid. - Mandatory Mandatory Mandatory
Biosafety Comply with the following descriptions except when complying with Biosafety in Attached Table 9-1 No.2. - - - -
Manage and keep record about use and discharge of laboratory animal. Recommended Mandatory Mandatory Mandatory
Waste Treatment Comply with the following descriptions except when complying with Waste Treatment in Attached Table 9-1 No.2. - - - -
Remove biological activation from litter after placing it in sealable container, etc., before disposal. Recommended Mandatory Mandatory Mandatory