Location of and Access to Laboratory |
Greenhouse (experiment zone): separation (division) from general experiment zone. Sealed zone (equipment storage area, experiment zone) inLevel 3·4 facility: separation from general zone |
Recommended |
Recommended |
Mandatory |
Mandatory |
Installation of storage place to keep personal clothing and laboratory suit before access to laboratory |
Recommended |
Recommended |
Mandatory |
Mandatory |
Installation of locking device in door (card, fingerprint recognition system, security system, etc.) |
Recommended |
Mandatory |
Mandatory |
Mandatory |
Installation of device that prevents interactive opening of doors in the zone (manual control allowed) |
- |
- |
Mandatory |
Mandatory |
Door: installation of door that enables air inflation or compression seal |
- |
- |
Recommended |
Mandatory |
Experiment Zone |
Greenhouse floor: absorptive or porous floor |
Mandatory |
Mandatory |
- |
- |
Facility interior: use of material in inner wall, ceiling, floor, etc.that is seamless and enables chemical sterilization and fumigation (full sealing of inner tube for Level 4 research facility) |
- |
- |
Mandatory |
Mandatory |
Installation of equipment for catching discharged water |
- |
Recommended |
Mandatory |
Mandatory |
Use of standard greenhouse glass or plastic material |
Mandatory |
Mandatory |
- |
- |
Full sealing by using enhanced plastic glass |
- |
- |
Mandatory |
Mandatory |
Use of screen window of 30mesh size |
Mandatory |
- |
- |
- |
Use of screen window of 30mesh size or over |
- |
Mandatory |
- |
- |
Prohibition on use of screen window and window |
- |
- |
Mandatory |
Mandatory |
Air-conditioning |
Air inside laboratory: maintenance of negative pressure and prevention of recirculation |
- |
- |
Mandatory |
Mandatory |
Ventilation system for independent air supply and discharge |
- |
- |
Mandatory |
Mandatory |
Installation of ventilation fan applied with back draft damper |
- |
- |
Mandatory |
Mandatory |
Installation of HEPA filter in air supply duct |
- |
- |
Recommended |
Mandatory |
Installation of HEPA filter in exhaust duct (installation of two-layer HEPA filter for Level 4 researchfacility) |
- |
- |
Mandatory |
Mandatory |
Installation of exhaust facility on roof or side |
Mandatory |
Mandatory |
- |
- |
Experimenter Safety Protection |
Installation of hand sterilizer and eye washer in laboratory |
- |
Recommended |
Mandatory |
Mandatory |
Installation of emergency shower facility in close location to dressing zone for taking off contaminated laboratory suit |
- |
- |
Mandatory |
Mandatory |
Installation of chemical shower equipment for undressing of contaminated laboratory suit |
- |
- |
- |
Mandatory |
Experiment Equipment |
Installation of autoclave |
- |
Mandatory |
Mandatory |
Mandatory |
No seam in work cabinet. Use of chemical resistance material in work cabinet |
- |
- |
Mandatory |
Mandatory |
Isolated cultivation, if needed, using closed-type plant cultivation equipment (e.g., bottle, plastic box) |
- |
Recommended |
Mandatory |
Mandatory |
Drainage and Waste Treatment |
Solid waste: installation of equipment that can remove biological activation, such as autoclave or chemical processing |
Recommended |
Mandatory |
Mandatory |
Mandatory |
Waste water from experiment: installation of equipment that can remove biological activation, such as autoclave or chemical processing |
- |
Recommended |
Mandatory |
Mandatory |
Exhaust through HEPA filter (two-layer HEPA filter processing for Level 4 research facility) |
- |
Recommended |
Mandatory |
Mandatory |
Catchment of discharged water: removal of pollution before discharge |
- |
- |
Mandatory |
Mandatory |
Installation of back draft prevention device for pipe |
- |
- |
- |
Mandatory |
Other Facilities |
Compliance with Other Facilitiesin Attached Table 9-1 No.1 |
|
|
|
|